Mella, HernánMura, JoaquínSotelo, JulioUribe, Sergio2022-11-302022-11-302021http://repositoriobibliotecas.uv.cl/handle/uvscl/7421We addressed comprehensively the performance of Shortest-Path HARP Refinement (SP-HR), SinMod, and DENSEanalysis using 2D slices of synthetic CSPAMM and DENSE images with realistic contrasts obtained from 3D phantoms. The three motion estimation techniques were interrogated under ideal and no-ideal conditions (with MR induced artifacts, noise, and through-plane motion), considering several resolutions and noise levels. Under noisy conditions, and for isotropic pixel sizes of 1.5 mm and 3.0 mm in CSPAMM and DENSE images respectively, the nRMSE obtained for the circumferential and radial strain components were 10.7 ± 10.8% and 25.5 ± 14.8% using SP-HR, 11.9 ± 2.5% and 29.3 ± 6.5% using SinMod, and 6.4 ± 2.0% and 18.2 ± 4.6% using DENSEanalysis. Overall, the results showed that SP-HR tends to fail for large tissue motions, whereas SinMod and DENSEanalysis gave accurate displacement and strain field estimations, being the last which performed the best.© 2021 Elsevier Inc. All rights reserved.TAGGING MRICSPAMMDENSE MRICARDIAC MRICARDIAC STRAINHARPSP-HRSINMODDENSEANALYSISA comprehensive comparison between shortest-path HARP refinement, SinMod, and DENSEanalysis processing tools applied to CSPAMM and DENSE imagesArticulohttps://doi.org/10.1016/j.mri.2021.07.001